Abbott Accent MRI
Approved changes feed: RSS · Atom
cpe:2.3:h:abbott:accent_mri:-:*:*:*:*:*:*:*
part: h version: - update: *
| Vendor | Abbott (175b6c33-e87a-5f0f-bbf2-691917441ae6) |
|---|---|
| Product | Accent Mri (d8691933-1303-53b9-bedb-13c257f3d6df) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from NVD CPE 2.0 feed |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2017-12716 |
not_vulnerable | 2026-06-03 14:36:36.600840 |
Details available
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Published: 2018-04-25T13:00:00.000Z
Updated: 2024-09-17T01:55:36.387Z |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2017-12714 |
not_vulnerable | 2026-06-03 14:36:36.596988 |
Details available
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Published: 2018-04-25T13:00:00.000Z
Updated: 2024-09-17T00:16:50.163Z |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2017-12712 |
not_vulnerable | 2026-06-03 14:36:36.589355 |
Details available
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Published: 2018-04-25T13:00:00.000Z
Updated: 2024-09-17T03:48:37.996Z |
Imported from gcve-enriched-dumps CVE data |