Medtronic Diabetes 751 MiniMed 530g Firmware
Approved changes feed: RSS · Atom
cpe:2.3:o:medtronicdiabetes:751_minimed_530g_firmware:-:*:*:*:*:*:*:*
part: o version: - update: *
| Vendor | Medtronicdiabetes (38bb25d6-78b6-5201-9bd6-9f68ddb675d6) |
|---|---|
| Product | 751 Minimed 530G Firmware (7d8bb1c6-8f19-5bb0-9571-00f9a2d64758) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from NVD CPE 2.0 feed |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2018-14781 |
vulnerable | 2026-06-03 14:38:12.590417 |
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
MEDIUM (5.3)
Medtronic MiniMed MMT
devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Published: 2018-08-13T22:00:00.000Z
Updated: 2025-05-22T16:33:08.385Z |
Imported from gcve-enriched-dumps CVE data |
Contribute
You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.