Approved changes feed: RSS · Atom

cpe:2.3:o:hospira:lifecare_pcainfusion_firmware:*:*:*:*:*:*:*:*

part: o version: * update: *

VendorHospira (c0d9e8bb-f70a-5b1e-b62f-95531b009d1a)
ProductLifecare Pcainfusion Firmware (788cd20e-bd86-5496-9cd3-05a0fb5c0180)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2015-3958 vulnerable 2026-06-08 05:06:38.883859 Details available
Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets.
Published: 2015-07-06T19:10:00.000Z
Updated: 2024-08-06T06:04:01.006Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2015-3957 vulnerable 2026-06-08 05:06:38.883514 Details available
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.
Published: 2015-07-06T19:10:00.000Z
Updated: 2024-08-06T06:04:01.150Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2015-3955 vulnerable 2026-06-08 05:06:38.882492 Details available
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.
Published: 2015-07-06T19:10:00.000Z
Updated: 2024-08-06T06:04:01.161Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2015-3459 vulnerable 2026-06-08 05:06:38.365018 Details available
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
Published: 2015-04-29T23:00:00.000Z
Updated: 2024-08-06T05:47:57.892Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2015-1011 vulnerable 2026-06-08 05:06:24.816304 Details available
Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors.
Published: 2015-07-06T19:10:00.000Z
Updated: 2024-08-06T04:26:11.531Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2014-5406 vulnerable 2026-06-08 05:05:47.841239 Hospira LifeCare PCA Infusion System
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Published: 2015-07-06T19:10:00.000Z
Updated: 2025-11-03T18:34:36.324Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.