Approved changes feed: RSS · Atom

cpe:2.3:a:hospira:plum_a+3_infusion_system:*:*:*:*:*:*:*:*

part: a version: * update: *

VendorHospira (c0d9e8bb-f70a-5b1e-b62f-95531b009d1a)
ProductPlum A+3 Infusion System (2e7cf312-045e-5813-af62-4a29ecc3423a)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2015-3956 vulnerable 2026-06-08 05:06:38.883089 Details available
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published: 2019-03-25T17:44:44.000Z
Updated: 2024-08-06T06:04:02.927Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2015-3954 vulnerable 2026-06-08 05:06:38.882054 Details available
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published: 2019-03-25T16:12:01.000Z
Updated: 2024-08-06T06:04:01.128Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2015-3953 vulnerable 2026-06-08 05:06:38.881536 Details available
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published: 2019-03-25T16:02:25.000Z
Updated: 2024-08-06T06:04:00.957Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2015-3952 vulnerable 2026-06-08 05:06:38.877061 Details available
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published: 2019-03-25T15:42:39.000Z