Approved changes feed: RSS · Atom

cpe:2.3:a:abbott_laboratories:accent_and_anthem:*:*:*:*:*:*:*:*

part: a version: * update: *

VendorAbbott Laboratories (04b5d773-4ef7-5224-bbcb-ac3acdda81f1)
ProductAccent And Anthem (8558616a-f0fb-5c2c-842d-0d55e783bece)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2017-12716 vulnerable 2026-06-03 14:36:36.600678 Details available
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Published: 2018-04-25T13:00:00.000Z
Updated: 2024-09-17T01:55:36.387Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.