Mycarelink 24950 Patient Monitor
Approved changes feed: RSS · Atom
cpe:2.3:h:medtronic:mycarelink_24950_patient_monitor:-:*:*:*:*:*:*:*
part: h version: - update: *
| Vendor | Medtronic (39ade870-5d3b-575a-9a52-512cca95c641) |
|---|---|
| Product | Mycarelink 24950 Patient Monitor (7b55abcc-594f-5e68-99bb-a750300ea29e) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from gcve-enriched-dumps CVE data |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2018-10626 |
not_vulnerable | 2026-06-08 05:10:35.214167 |
Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity
MEDIUM (4.4)
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An
attacker who obtains per-product credentials from the monitor and paired
implantable cardiac device information can potentially upload invalid
data to the Medtronic CareLink network.
Published: 2018-08-10T18:00:00.000Z
Updated: 2026-05-19T14:32:21.570Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2018-10622 |
not_vulnerable | 2026-06-08 05:10:35.209237 |
Medtronic MyCareLink Patient Monitor Network Credential Weakness
MEDIUM (5.2)
Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.
Published: 2018-08-10T18:00:00.000Z
Updated: 2026-06-22T20:42:16.327Z Reference links |
Imported from gcve-enriched-dumps CVE data |
Contribute
You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.