cpe:2.3:a:ascensia:contour_diabetes:*:*:*:*:*:android:*:*

part: a version: * update: *

VendorAscensia (103597d2-e8bc-52e2-8e79-ba6af1430495)
ProductContour Diabetes (f867962e-867f-5c17-aa2c-4b102104ead5)
Edition*
Language*
Software edition*
Target softwareandroid
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2018-18979 vulnerable 2026-06-08 05:11:15.035168 Details available
An issue was discovered in the Ascensia Contour NEXT ONE application for Android before 2019-01-15. It has a statically coded initialization vector. Extraction of the initialization vector is necessary for deciphering communications between this application and the backend server. This, in combination with retrieving any user's encrypted data from the Ascensia cloud through another vulnerability, allows an attacker to obtain and modify any patient's medical information.
Published: 2019-05-06T19:19:20.000Z
Updated: 2024-08-05T11:23:08.805Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-18978 vulnerable 2026-06-08 05:11:15.034866 Details available
An issue was discovered in the Ascensia Contour NEXT ONE application for Android before 2019-01-15. It has a statically coded encryption key. Extraction of the encryption key is necessary for deciphering communications between this application and the backend server. This, in combination with retrieving any user's encrypted data from the Ascensia cloud through another vulnerability, allows an attacker to obtain and modify any patient's medical information.
Published: 2019-05-06T19:18:26.000Z
Updated: 2024-08-05T11:23:08.583Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-18977 vulnerable 2026-06-08 05:11:15.034444 Details available
An issue was discovered in the Ascensia Contour NEXT ONE application for Android before 2019-01-15. An attacker may reverse engineer the codebase to extract sensitive data that contributes to the disclosure of medical information of patients utilizing the Ascensia platform. This occurs because of weak obfuscation.
Published: 2019-05-06T19:14:58.000Z
Updated: 2024-08-05T11:23:08.914Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-18976 vulnerable 2026-06-08 05:11:15.033936 Details available
An issue was discovered in the Ascensia Contour NEXT ONE application for iOS and Android before 2019-01-15. An attacker may retrieve encrypted medical information of any user of the Ascensia cloud platform by performing Direct Object References with a series of user ID values. (This information can be decrypted through a different vulnerability.)
Published: 2019-05-06T19:12:22.000Z
Updated: 2024-08-05T11:23:08.568Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.