Approved changes feed: RSS · Atom

cpe:2.3:a:medtronic:carelink_2090_programmer:*:*:*:*:*:*:*:*

part: a version: * update: *

VendorMedtronic (39ade870-5d3b-575a-9a52-512cca95c641)
ProductCarelink 2090 Programmer (3d5c8002-9ec4-5321-b9a7-ed5f32c9a576)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2019-6540 vulnerable 2026-06-08 05:14:11.920757 Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
MEDIUM (6.5)
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Published: 2019-03-26T17:47:23.000Z
Updated: 2025-05-22T19:23:42.770Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2019-6538 vulnerable 2026-06-08 05:14:11.883957 Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control
CRITICAL (9.3)
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Published: 2019-03-25T21:26:03.000Z
Updated: 2025-05-22T19:26:15.126Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-18984 vulnerable 2026-06-08 05:11:15.043963 Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers Missing Encryption of Sensitive Data
MEDIUM (4.6)
Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .
Published: 2018-12-14T15:00:00.000Z
Updated: 2025-05-22T16:40:53.366Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute