Patient Information Center Ix (Picix)
Approved changes feed: RSS · Atom
cpe:2.3:a:philips:patient_information_center_ix_(picix):*:*:*:*:*:*:*:*
part: a version: * update: *
| Vendor | Philips (b8a9f1cf-d82f-5e30-aade-8a8018bb7a6f) |
|---|---|
| Product | Patient Information Center Ix (Picix) (e961ba99-2283-5a18-b74b-6fc3a9ca45db) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from gcve-enriched-dumps CVE data |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2020-16228 |
vulnerable | 2026-06-08 05:19:27.754467 |
Philips Patient Monitoring Devices Improper Check for Certificate Revocation
In Patient Information Center iX (PICiX) Versions C.02 and C.03,
PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors
MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and
prior, the software does not check or incorrectly checks the revocation
status of a certificate, which may cause it to use a compromised
certificate.
Published: 2020-09-11T12:58:13.000Z
Updated: 2024-08-04T13:37:54.179Z |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-16224 |
vulnerable | 2026-06-08 05:19:27.588824 |
Philips Patient Monitoring Devices Improper Handling of Length Parameter Inconsistency
In Patient Information Center iX (PICiX) Versions C.02, C.03, the
software parses a formatted message or structure but does not handle or
incorrectly handles a length field that is inconsistent with the actual
length of the associated data, causing the application on the
surveillance station to restart.
Published: 2020-09-11T13:05:19.000Z
Updated: 2024-08-04T13:37:53.948Z |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-16222 |
vulnerable | 2026-06-08 05:19:27.587963 |
Philips Patient Monitoring Devices Improper Authentication
In Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and
PerformanceBridge Focal Point Version A.01, when an actor claims to have
a given identity, the software does not prove or insufficiently proves
the claim is correct.
Published: 2020-09-11T12:55:23.000Z
Updated: 2024-08-04T13:37:54.092Z |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-16220 |
vulnerable | 2026-06-08 05:19:27.585895 |
Philips Patient Monitoring Devices Improper Validation of Syntactic Correctness of Input
In Patient Information Center iX (PICiX) Versions C.02, C.03,
PerformanceBridge Focal Point Version A.01, the product receives input
that is expected to be well-formed (i.e., to comply with a certain
syntax) but it does not validate or incorrectly validates that the input
complies with the syntax, causing the certificate enrollment service to
crash. It does not impact monitoring but prevents new devices from
enrolling.
Published: 2020-09-11T13:11:22.000Z
Updated: 2024-08-04T13:37:53.493Z |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-16218 |
vulnerable | 2026-06-08 05:19:27.582521 |
Philips Patient Monitoring Devices Cross-site Scripting
In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the
software does not neutralize or incorrectly neutralizes
user-controllable input before it is placed in output that is then used
as a webpage and served to other users. Successful exploitation could
lead to unauthorized access to patient data via a read-only web
application.
Published: 2020-09-11T12:50:08.000Z
Updated: 2024-08-04T13:37:54.157Z |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-16214 |
vulnerable | 2026-06-08 05:19:27.559039 | db.gcve.eu details were skipped to keep the page responsive. | Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-16212 |
vulnerable | 2026-06-08 05:19:27.556142 | db.gcve.eu details were skipped to keep the page responsive. | Imported from gcve-enriched-dumps CVE data |
Contribute
You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.