Approved changes feed: RSS · Atom

cpe:2.3:o:medtronic:mycarelink_smart_model_25000_firmware:*:*:*:*:*:*:*:*

part: o version: * update: *

VendorMedtronic (39ade870-5d3b-575a-9a52-512cca95c641)
ProductMycarelink Smart Model 25000 Firmware (1c6cb774-0302-50d1-97a2-00db2a1a145f)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2020-27252 vulnerable 2026-06-08 05:23:52.466563 Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition
HIGH (8.8)
Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
Published: 2020-12-14T19:19:00.000Z
Updated: 2025-05-22T19:39:10.747Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2020-25187 vulnerable 2026-06-08 05:22:36.395483 Medtronic MyCareLink Smart Heap-based Buffer Overflow
HIGH (8.8)
Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
Published: 2020-12-14T19:18:52.000Z
Updated: 2025-05-22T19:37:12.795Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2020-25183 vulnerable 2026-06-08 05:22:36.383100 Medtronic MyCareLink Smart Improper Authentication
HIGH (8)
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
Published: 2020-12-14T19:18:44.000Z
Updated: 2025-05-22T19:34:29.069Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.