Zoom Latitude Pogrammer/Recorder/Monitor 3120
Approved changes feed: RSS · Atom
cpe:2.3:h:bostonscientific:zoom_latitude_pogrammer\/recorder\/monitor_3120:-:*:*:*:*:*:*:*
part: h version: - update: *
| Vendor | Bostonscientific (c8df7a5b-7e1e-5217-898a-6bcec9aa1ae6) |
|---|---|
| Product | Zoom Latitude Pogrammer/Recorder/Monitor 3120 (f685852a-5021-5e13-81ae-17140ccb20ad) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from gcve-enriched-dumps CVE data |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2021-38400 |
not_vulnerable | 2026-06-08 05:32:56.021447 |
Use of Password Hash with Insufficient Computational Effort for Boston Scientific Zoom Latitude
MEDIUM (6.9)
An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password.
Published: 2021-10-04T17:35:28.144Z
Updated: 2024-09-16T23:31:04.619Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2021-38398 |
not_vulnerable | 2026-06-08 05:32:56.020291 |
Reliance on Component that is not Updateable for Boston Scientific Zoom Latitude
MEDIUM (6.5)
The affected device uses off-the-shelf software components that contain unpatched vulnerabilities. A malicious attacker with physical access to the affected device could exploit these vulnerabilities.
Published: 2021-10-04T17:35:06.578Z
Updated: 2024-09-16T17:58:22.688Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2021-38396 |
not_vulnerable | 2026-06-08 05:32:56.015985 |
Missing Support Integrity Check for Boston Scientific Zoom Latitude
MEDIUM (6.5)
The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.
Published: 2021-10-04T17:35:21.072Z
Updated: 2024-09-16T20:46:57.503Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2021-38394 |
not_vulnerable | 2026-06-08 05:32:56.007405 |
Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston Scientific Zoom Latitude
MEDIUM (6.2)
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.
Published: 2021-10-04T17:34:58.296Z
Updated: 2024-09-16T18:59:19.736Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2021-38392 |
not_vulnerable | 2026-06-08 05:32:56.006457 |
Improper Access Control for Boston Scientific Zoom Latitude
MEDIUM (6.5)
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
Published: 2021-10-04T17:35:13.879Z
Updated: 2024-09-17T03:33:08.540Z Reference links |
Imported from gcve-enriched-dumps CVE data |
Contribute
You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.