Approved changes feed: RSS · Atom

cpe:2.3:a:contec_health:cms8000_contec_icu_ccu_vital_signs_patient_monitor:*:*:*:*:*:*:*:*

part: a version: * update: *

VendorContec Health (76fe6a44-c80e-5177-bf72-cffb907d8cb8)
ProductCms8000 Contec Icu Ccu Vital Signs Patient Monitor (58e206b2-62e6-5146-866c-3342f80e43d5)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2022-3027 vulnerable 2026-06-08 05:47:19.065147 Contec Health CMS8000
MEDIUM (5.7)
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
Published: 2022-09-13T14:54:58.631Z
Updated: 2025-04-16T16:10:02.402Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2022-38453 vulnerable 2026-06-08 05:47:16.413463 Contec Health CMS8000
LOW (3)
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.
Published: 2022-09-13T14:55:03.842Z
Updated: 2025-04-16T16:09:53.917Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2022-38100 vulnerable 2026-06-08 05:47:15.720968 Contec Health CMS8000
HIGH (7.5)
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Published: 2022-09-13T14:54:50.356Z
Updated: 2025-04-16T16:10:19.613Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2022-38069 vulnerable 2026-06-08 05:47:14.525163 Contec Health CMS8000
MEDIUM (4.3)
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
Published: 2022-09-13T14:54:54.113Z
Updated: 2025-04-16T16:10:11.963Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2022-36385 vulnerable 2026-06-08 05:46:07.227481 Contec Health CMS8000
MEDIUM (6.8)
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
Published: 2022-09-13T14:54:35.692Z
Updated: 2025-04-16T17:47:37.455Z
Reference links