Approved changes feed: RSS · Atom

cpe:2.3:h:roche:cobas_h_232:-:*:*:*:*:*:*:*

part: h version: - update: *

VendorRoche (5c2ccb88-3eeb-5293-a5a4-1a379f643347)
ProductCobas H 232 (085c645d-59ac-5403-9cf7-8be5205609dc)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from NVD CPE 2.0 feed

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2018-18565 not_vulnerable 2026-06-03 14:38:28.024007 Details available
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.
Published: 2018-11-20T19:00:00.000Z
Updated: 2024-08-05T11:15:59.675Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-18564 not_vulnerable 2026-06-03 14:38:28.023466 Details available
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). Improper access control allows attackers in the adjacent network to change the instrument configuration.
Published: 2018-11-20T19:00:00.000Z
Updated: 2024-08-05T11:15:59.689Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-18563 not_vulnerable 2026-06-03 14:38:28.017000 Details available
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.
Published: 2018-11-20T19:00:00.000Z
Updated: 2024-08-05T11:16:00.005Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-18562 not_vulnerable 2026-06-03 14:38:28.016432 Details available
An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Unit before 03.01.04. Weak access credentials may enable attackers in the adjacent network to gain unauthorized service access via a service interface.
Published: 2018-11-20T19:00:00.000Z
Updated: 2024-08-05T11:15:59.590Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-18561 not_vulnerable 2026-06-03 14:38:28.012409 Details available
An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Unit before 03.01.04. Insecure permissions in a service interface may allow authenticated attackers in the adjacent network to execute arbitrary commands on the operating system.
Published: 2018-11-20T19:00:00.000Z
Updated: 2024-08-05T11:15:59.870Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.