Approved changes feed: RSS · Atom

cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*

part: o version: * update: *

VendorBaxter (a179c94a-1aed-5d11-8395-fe3753820594)
ProductLife2000 Ventilator Firmware (25b5676a-4b78-5f65-a3ae-7ffff6ea68d9)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2024-9834 vulnerable 2026-06-08 07:00:28.749095 Improper data protection on Life2000 ventilator serial interface
CRITICAL (9.3)
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Published: 2024-11-14T20:57:22.734Z
Updated: 2024-11-15T15:10:40.157Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2024-9832 vulnerable 2026-06-08 07:00:28.747918 No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
CRITICAL (9.3)
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Published: 2024-11-14T21:03:16.721Z
Updated: 2024-11-18T15:37:00.311Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2024-48974 vulnerable 2026-06-08 06:50:11.784887 Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
CRITICAL (9.3)
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Published: 2024-11-14T21:27:10.284Z
Updated: 2024-11-18T15:32:38.488Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2024-48973 vulnerable 2026-06-08 06:50:11.784386 Debug port on Life2000 Ventilator serial interface is enabled by default
CRITICAL (9.3)
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Published: 2024-11-14T21:24:14.099Z
Updated: 2024-11-18T15:33:44.924Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2024-48971 vulnerable 2026-06-08 06:50:11.783866 Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
CRITICAL (9.3)
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
Published: 2024-11-14T21:13:36.036Z
Updated: 2024-11-15T21:06:24.325Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2024-48970 vulnerable 2026-06-08 06:50:11.783341 db.gcve.eu details were skipped to keep the page responsive. Imported from gcve-enriched-dumps CVE data
CVE:CVE-2024-48967 vulnerable 2026-06-08 06:50:11.782785 db.gcve.eu details were skipped to keep the page responsive. Imported from gcve-enriched-dumps CVE data
CVE:CVE-2024-48966 vulnerable 2026-06-08 06:50:11.781198 db.gcve.eu details were skipped to keep the page responsive. Imported from gcve-enriched-dumps CVE data

Contribute