Approved changes feed: RSS · Atom

cpe:2.3:a:philips:intellispace_cardiovascular_(iscv):*:*:*:*:*:*:*:*

part: a version: * update: *

VendorPhilips (b8a9f1cf-d82f-5e30-aade-8a8018bb7a6f)
ProductIntellispace Cardiovascular (Iscv) (a64a8265-9751-5baa-b171-40597b5731c5)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2025-2230 vulnerable 2026-06-08 07:16:56.521696 Philips Intellispace Cardiovascular (ISCV) Improper Authentication
HIGH (7.7)
A flaw exists in the Windows login flow where an AuthContext token can be exploited for replay attacks and authentication bypass.
Published: 2025-03-13T18:14:43.573Z
Updated: 2025-03-17T20:05:56.939Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2025-2229 vulnerable 2026-06-08 07:16:56.519137 Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials
HIGH (7.7)
A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.
Published: 2025-03-13T18:17:31.256Z
Updated: 2025-03-17T20:04:56.100Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.