Approved changes feed: RSS · Atom

cpe:2.3:a:eppendorf:bioflo_320:*:*:*:*:*:*:*:*

part: a version: * update: *

VendorEppendorf (485e96ee-f444-5a39-8cb4-e0c6eb44e6f5)
ProductBioflo 320 (78b3f018-8936-5dea-ab58-8189cbbb124c)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*

Notes

Imported from gcve-enriched-dumps CVE data

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2026-7251 vulnerable 2026-06-08 08:08:56.694165 Eppendorf BioFlo 320 Use of hard-coded password
CRITICAL (9.8)
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
Published: 2026-05-26T17:06:46.563Z
Updated: 2026-06-04T21:10:51.687Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.

Inspect record JSON

Open JSON directly

Open this panel to browse, search, copy or download the JSON record.