GE Healthcare Clinical Information Center MP100D Firmware 5.0
Approved changes feed: RSS · Atom
cpe:2.3:o:gehealthcare:clinical_information_center_mp100d_firmware:5.0:*:*:*:*:*:*:*
part: o version: 5.0 update: *
| Vendor | Gehealthcare (49dacc94-769c-5b2a-9768-4b15fd7cb9d8) |
|---|---|
| Product | Clinical Information Center Mp100D Firmware (c4a95d5a-8606-5284-b19a-22cd66ea118b) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from NVD CPE 2.0 feed |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2020-6966 |
vulnerable | 2026-06-03 14:43:04.658727 |
Details available
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilize a weak encryption scheme for remote desktop control, which may allow an attacker to obtain remote code execution of devices on the network.
Published: 2020-01-24T17:07:00.000Z
Updated: 2024-08-04T09:18:02.517Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-6965 |
vulnerable | 2026-06-03 14:43:04.657782 |
Details available
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, a vulnerability in the software update mechanism allows an authenticated attacker to upload arbitrary files on the system through a crafted update package.
Published: 2020-01-24T17:06:53.000Z
Updated: 2024-08-04T09:18:02.543Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-6964 |
vulnerable | 2026-06-03 14:43:04.656310 |
Details available
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.
Published: 2020-01-24T16:32:13.000Z
Updated: 2024-08-04T09:18:02.496Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-6963 |
vulnerable | 2026-06-03 14:43:04.655274 |
Details available
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
Published: 2020-01-24T16:31:37.000Z
Updated: 2024-08-04T09:18:01.475Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-6962 |
vulnerable | 2026-06-03 14:43:04.651090 |
Details available
In ApexPro Telemetry Server, Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Telemetry Server Version 4.3, CARESCAPE Central Station (CSCS) Versions 1.X CARESCAPE Central Station (CSCS) Versions 2.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, an input validation vulnerability exists in the web-based system configuration utility that could allow an attacker to obtain arbitrary remote code execution.
Published: 2020-01-24T16:07:54.000Z
Updated: 2024-08-04T09:18:01.628Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-6961 |
vulnerable | 2026-06-03 14:43:04.634041 |
Details available
In ApexPro Telemetry Server, Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Telemetry Server Version 4.3, CARESCAPE Central Station (CSCS) Versions 1.X, a vulnerability exists in the affected products that could allow an attacker to obtain access to the SSH private key in configuration files.
Published: 2020-01-24T16:07:45.000Z
Updated: 2024-08-04T09:18:02.493Z Reference links |
Imported from gcve-enriched-dumps CVE data |
Contribute
You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.