Approved changes feed: RSS · Atom

cpe:2.3:o:gehealthcare:clinical_information_center_mp100r_firmware:5.0:*:*:*:*:*:*:*

part: o version: 5.0 update: *

VendorGehealthcare (49dacc94-769c-5b2a-9768-4b15fd7cb9d8)
ProductClinical Information Center Mp100R Firmware (de44f1e3-2e29-5dd4-a748-23e1b9b040f4)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from NVD CPE 2.0 feed

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2020-6966 vulnerable 2026-06-08 05:27:11.338549 Details available
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilize a weak encryption scheme for remote desktop control, which may allow an attacker to obtain remote code execution of devices on the network.
Published: 2020-01-24T17:07:00.000Z
Updated: 2024-08-04T09:18:02.517Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2020-6965 vulnerable 2026-06-08 05:27:11.333601 Details available
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, a vulnerability in the software update mechanism allows an authenticated attacker to upload arbitrary files on the system through a crafted update package.
Published: 2020-01-24T17:06:53.000Z
Updated: 2024-08-04T09:18:02.543Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2020-6964 vulnerable 2026-06-08 05:27:11.331733 Details available
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.
Published: 2020-01-24T16:32:13.000Z
Updated: 2024-08-04T09:18:02.496Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2020-6963 vulnerable 2026-06-08 05:27:11.330159 Details available
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
Published: 2020-01-24T16:31:37.000Z
Updated: 2024-08-04T09:18:01.475Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2020-6962 vulnerable 2026-06-08 05:27:11.326818 Details available
In ApexPro Telemetry Server, Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Telemetry Server Version 4.3, CARESCAPE Central Station (CSCS) Versions 1.X CARESCAPE Central Station (CSCS) Versions 2.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, an input validation vulnerability exists in the web-based system configuration utility that could allow an attacker to obtain arbitrary remote code execution.
Published: 2020-01-24T16:07:54.000Z
Updated: 2024-08-04T09:18:01.628Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2020-6961 vulnerable 2026-06-08 05:27:11.313287 db.gcve.eu details were skipped to keep the page responsive. Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.