Baxter Sigma Spectrum Infusion System Firmware 8.0
Approved changes feed: RSS · Atom
cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:*
part: o version: 8.0 update: *
| Vendor | Baxter (a179c94a-1aed-5d11-8395-fe3753820594) |
|---|---|
| Product | Sigma Spectrum Infusion System Firmware (36ab6c10-ae6d-5f1d-a6f6-70cf9def9a6b) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from NVD CPE 2.0 feed |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2020-12047 |
vulnerable | 2026-06-03 14:41:33.199399 |
Details available
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.
Published: 2020-06-29T13:41:37.000Z
Updated: 2024-08-04T11:48:57.938Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-12045 |
vulnerable | 2026-06-03 14:41:33.198666 |
Details available
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.
Published: 2020-06-29T13:43:40.000Z
Updated: 2024-08-04T11:48:58.009Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-12043 |
vulnerable | 2026-06-03 14:41:33.198206 |
Details available
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted.
Published: 2020-06-29T13:41:42.000Z
Updated: 2024-08-04T11:48:57.976Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-12041 |
vulnerable | 2026-06-03 14:41:33.192416 |
Details available
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
Published: 2020-06-29T13:43:52.000Z
Updated: 2024-08-04T11:48:57.710Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-12040 |
vulnerable | 2026-06-03 14:41:33.192046 |
Details available
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Published: 2020-06-29T13:45:58.000Z
Updated: 2024-08-04T11:48:57.530Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2020-12039 |
vulnerable | 2026-06-03 14:41:33.190918 |
Details available
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.