Drtrust Electrocardiogram Pen Firmware 2.00.08
Approved changes feed: RSS · Atom
cpe:2.3:o:drtrust:electrocardiogram_pen_firmware:2.00.08:*:*:*:*:*:*:*
part: o version: 2.00.08 update: *
| Vendor | Drtrust (f38cca17-d2d2-589c-be1d-f55ae0f8a845) |
|---|---|
| Product | Electrocardiogram Pen Firmware (81c5854b-53a5-5ced-a3ff-7e2e429dbed6) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from NVD CPE 2.0 feed |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2020-15486 |
vulnerable | 2026-06-08 05:19:25.972732 |
Details available
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Published: 2020-08-26T15:14:59.000Z
Updated: 2024-08-04T13:15:20.806Z |
Imported from gcve-enriched-dumps CVE data |
Contribute
You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.