cpe:2.3:o:philips:intellivue_mx100_firmware:-:*:*:*:*:*:*:*

part: o version: - update: *

VendorPhilips (b8a9f1cf-d82f-5e30-aade-8a8018bb7a6f)
ProductIntellivue Mx100 Firmware (4890fff9-9802-5806-8960-186ce7f455f2)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from NVD CPE 2.0 feed

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2020-16228 vulnerable 2026-06-03 14:41:47.568412 Philips Patient Monitoring Devices Improper Check for Certificate Revocation
In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior, the software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate.
Published: 2020-09-11T12:58:13.000Z
Updated: 2024-08-04T13:37:54.179Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2020-16216 vulnerable 2026-06-03 14:41:47.341988 Philips Patient Monitoring Devices Improper Input Validation
In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly, which can induce a denial-of-service condition through a system restart.
Published: 2020-09-11T13:06:55.000Z
Updated: 2024-08-04T13:37:54.166Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-10601 vulnerable 2026-06-03 14:37:58.468670 Details available
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
Published: 2018-06-05T20:00:00.000Z
Updated: 2024-09-17T02:36:04.782Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-10599 vulnerable 2026-06-03 14:37:58.462317 Details available
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet.
Published: 2018-06-05T20:00:00.000Z
Updated: 2024-09-16T16:33:25.211Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-10597 vulnerable 2026-06-03 14:37:58.452860 Details available
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory ("write-what-where") from an attacker-chosen device address within the same subnet.