Approved changes feed: RSS · Atom

cpe:2.3:h:philips:intellivue_np90:-:*:*:*:*:*:*:*

part: h version: - update: *

VendorPhilips (b8a9f1cf-d82f-5e30-aade-8a8018bb7a6f)
ProductIntellivue Np90 (f2bb06e6-8c9e-5ef2-98c0-33dda2e334e0)
Edition*
Language*
Software edition*
Target software*
Target hardware*
Other*
NotesImported from NVD CPE 2.0 feed

PURL mappings

PURLSourceLast updated
No PURL mappings for this CPE yet.

Vulnerability references

IdentifiercpeApplicabilitySubmitteddb.gcve.eu detailsRationale
CVE:CVE-2018-10601 not_vulnerable 2026-06-08 05:10:35.157155 Details available
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
Published: 2018-06-05T20:00:00.000Z
Updated: 2024-09-17T02:36:04.782Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-10599 not_vulnerable 2026-06-08 05:10:35.151594 Details available
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet.
Published: 2018-06-05T20:00:00.000Z
Updated: 2024-09-16T16:33:25.211Z
Reference links
Imported from gcve-enriched-dumps CVE data
CVE:CVE-2018-10597 not_vulnerable 2026-06-08 05:10:35.131383 Details available
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory ("write-what-where") from an attacker-chosen device address within the same subnet.
Published: 2018-06-05T20:00:00.000Z
Updated: 2024-09-16T23:46:22.088Z
Reference links
Imported from gcve-enriched-dumps CVE data

Contribute

You can submit an edit proposal for this CPE entry or suggest a related product/vendor addition using the action button above.