CONTEC Health CMS8000 Firmware
Approved changes feed: RSS · Atom
cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*
part: o version: - update: *
| Vendor | Contechealth (3f9068b1-8e33-5e30-8206-6857788b4917) |
|---|---|
| Product | Cms8000 Firmware (e4c0637f-f6a6-599a-bc15-07acf12720c8) |
| Edition | * |
| Language | * |
| Software edition | * |
| Target software | * |
| Target hardware | * |
| Other | * |
| Notes | Imported from NVD CPE 2.0 feed |
PURL mappings
| PURL | Source | Last updated |
|---|---|---|
| No PURL mappings for this CPE yet. | ||
Vulnerability references
| Identifier | cpeApplicability | Submitted | db.gcve.eu details | Rationale |
|---|---|---|---|---|
CVE:CVE-2022-3027 |
vulnerable | 2026-06-08 05:47:19.065323 |
Contec Health CMS8000
MEDIUM (5.7)
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
Published: 2022-09-13T14:54:58.631Z
Updated: 2025-04-16T16:10:02.402Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2022-38453 |
vulnerable | 2026-06-08 05:47:16.413498 |
Contec Health CMS8000
LOW (3)
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.
Published: 2022-09-13T14:55:03.842Z
Updated: 2025-04-16T16:09:53.917Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2022-38100 |
vulnerable | 2026-06-08 05:47:15.721014 |
Contec Health CMS8000
HIGH (7.5)
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Published: 2022-09-13T14:54:50.356Z
Updated: 2025-04-16T16:10:19.613Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2022-38069 |
vulnerable | 2026-06-08 05:47:14.525972 |
Contec Health CMS8000
MEDIUM (4.3)
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
Published: 2022-09-13T14:54:54.113Z
Updated: 2025-04-16T16:10:11.963Z Reference links |
Imported from gcve-enriched-dumps CVE data |
CVE:CVE-2022-36385 |
vulnerable | 2026-06-08 05:46:07.228307 |
Contec Health CMS8000
MEDIUM (6.8)
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
Published: 2022-09-13T14:54:35.692Z
Updated: 2025-04-16T17:47:37.455Z Reference links |
Imported from gcve-enriched-dumps CVE data |
Contribute